Helicobacter Pylori antigen Hp-Ag uncut sheet

Short Description:

Helicobacter Pylori antigen Hp-Ag uncut sheet

Methodology: Latex 


  • Testing time: 15 minutes
  • Storage temperature: 2°C-30°C
  • Valid Time: 24 months
  • Accurancy: More than 99%
  • Specification: 25 test/box
  • Product Detail

    Product Tags

    1. INTEDED USE

    The H. pylori Antigen Rapid Test Device (the “uncut sheet” format for manufacturing use) is  intended for the qualitative detection of Helicobacter pylori (H. pylori) antigen in human stool specimens.

    This test is intended for in vitro diagnostic professional use and serves as an aid in the diagnosis of H. pylori infection and for monitoring the effectiveness of eradication therapy in patients presenting with clinical signs and symptoms suggestive of gastroduodenal disorders such as active chronic gastritis, gastric ulcer, or duodenal ulcer.

    The device provides presumptive (preliminary) screening test results. The test result should be interpreted in conjunction with patient history, clinical presentation, and other laboratory findings.

    This product is intended for use by trained healthcare professionals or laboratory personnel. This product is for manufacturing use only and is intended for further processing into final device format.

     

    2. PRODUCT SPECIFICATION

    Model No. HP-AG Uncut sheet
    Methodology Latex 
    Sample Type Faces 
    Time to Result 10-15mins
    Storage 2~30 ℃/36~86℉
    Shelf Life 24 months
    Certificate ISO13485, CE Certificate, UCKA MHRA Certificate
    uncut sheet
    uncut sheet

    Feature

      High sensitive

    •  result reading in 15-20 minutes

    •  Easy operation

    •  High Accuracy

    PACKING

    * 20 bags /CTN

    * Aluminum foil bag labeling

    * shrink wrap

     

    3.TEST METHOD

    1 Use sample collection tube to collect, mix well and dilute the sample. Use sampling stick to take about 30mg of stool. Then, transfer the stool to a sample collection tube containing sample diluent, tighten by rotating, and shake sufficiently.
    2 If stool of patient with diarrhea is loose, use disposable pipette to draw the sample, add 3 drops (about 100μL) of sample-to-sample collection tube, and shake the sample and sample diluent sufficiently.
    3 Take the test device out of the aluminum foil bag, place it on a horizontal worktable flat, and make a proper mark.
    4 Discard the first two drops of diluted sample. Then, vertically, and slowly add 3 drops (about 100μL) of bubble-Free diluted sample to the center of sample hole of the test device and start timing.
    5 The result shall be read within 10-15 minutes. Test result obtained after 15 minutes is invalid (for detail about result see Interpretation of Test Results).

     

    4.RESULT EVALUATION AND EXPLANATION

    bc1a84f6-7a47-4ea3-988b-cd569705d853

    5.CLINICAL PERFORMANCE

    Clinical performance of this product’s assessed through collection of 315 cases of clinical samples. Sample providers include patients infected with helicobacter pylori and normal subjects. Use marketed kit of turbidimetric inhibition immunoassays as reference reagent to compare with WIZ BIOTECH reagent.

    Baysen Result of hp--ag
    Test result of Reference reagent Positive coincidence rate:
    98.91%(95%C.I.96.12%~99.70%)
    Negative coincidence rate:
    100.00(95%C.I.97.15%~100.00%)
    Total coincidence rate:99.37% (95%C.I. 97.71%~99.83%)
    Positive  Negative  Total
    Positive 182 0 182
    Negative  2 131 133
    Total 184 131 315

    6. CERTIFICATE

    * ISO System Certificate

    * CE Certificate

    * EU Registration

    * UCKA MHRA Registration

    7.EXHIBITION

    exhibition